Validation of a fully autonomous phosphate analyser based on a microfluidic lab-on-a-chip
Slater, Conor and Cleary, John and Lau, King-Tong and Snakenborg, D. and Corcoran, Brian and Kutter, J.P. and Diamond, Dermot (2010) Validation of a fully autonomous phosphate analyser based on a microfluidic lab-on-a-chip. Water Science & Technology - WST, 61 (7). ISSN 0273-1223 Full text available as: AbstractThis work describes the design of a phosphate analyser that utilises a microfluidic lab-on-a-chip. The analyser contains all the required chemical storage, pumping and electronic components to carry out a complete phosphate assay. The system is self-calibrating and self-cleaning, thus capable of long-term operation. This was proven by a bench top calibration of the analyser using standard solutions and also by comparing the analyser's performance to a commercially available phosphate monitor installed at a waste water treatment plant. The output of the microfluidic lab-on-a-chip analyser was shown to have sensitivity and linear range equivalent to the commercially available monitor and also the ability to operate over an extended period of time. Download statistics